Designed for analysis + monitoring of “light” hydrocarbons.
It is used for the measurement of volatile hydrocarbon in the range
of C2 to C6 i.e. from Ethane through 1,3-Butadiene up to Benzene
(and other C6 compounds).
The instrument uses a 6-port valve with a single sorbent trap, and a metallic capillary column.
It has a minimum detection level as low as 15 ppt (for 1,3-Butadiene for e.g.) in ambient air, and analyses up to 32 compounds during a single analysis.
The airmo VOC C2 – C6 uses a thermoelectric cooling system to control the temperature of the sample trap when required, for example for the analysis of Ethyne.
Miniaturization, sensitivity, mobility and flexibility are its main features.
Everything from the sample port up to the data storage is integrated in a 19"- rack 4U.
Programmed gradient temperature of the oven and pressure/flow control of the carrier gas by the piezo-valve.
Before delivery the analyser is tested for one week by our quality control department.
The Vistachrom software enables the user to view and store data on a PC. It provides comfortable user friendly utilities to recalculate, calibrate and export data and to configure the measurement.
The software allows the calculation of retention time, area, mass or concentration profiles.